Product introduction:
The TVA2020C toxic volatile gas analyzer is the only portable on-site analyzer with the technology of flame ionization (FID) and photo ionization (PID) double detector simultaneously.
Detailed description of the product
FID/PID dual detector technology;
Light and compact design;
The operation of the field is simple;
There is no need for PC application software.
PDA hand operator
The Thermo Scientific TVA2020C toxic volatile gas analyzer has the ability to detect both organic and inorganic compounds at the same time. The TVA2020C analyzer can be applied to field repair detection, landfill site monitoring, and regular regional environmental survey, which is followed by the 21 EPA monitoring method.
TVA2020C is equipped with a highly sensitive flame ionization detector (FID) to measure the concentration of organic compounds. FID has a wide range of dynamic and linear measurement, stable response and good repeatability. The analyzer with both FID and PID dual detectors, which can work at the same time, has a stronger analytical ability. Compared with the instrument of single detector, dual detector analyzer can respond to all organic compounds and other inorganic compounds at the same time. Compared with the same volume of other instruments, it can provide more comprehensive gas coverage.
After correcting, TVA2020C can do some custom settings, such as setting up internal data recording parameters, uploading monitoring paths, building Bluetooth connections, setting alarm levels, selecting response factors, etc.
The optional Bluetooth communication function can send a lot of measurement data directly to the handheld device of the built-in LDAR software, so it doesn't need to be transmitted to the monitoring system after measurement, and it can also monitor the path information better.
TVA2020C reduces the weight of the earlier model by 21%, and is more compact than most single FID instruments. This portable, compact, portable design can really reduce the fatigue of the user. In addition, some of the TVA2020C configurations are optional, such as the basic / enhanced probe, the PDA hand operator, the portable box and the hydrogen gas valve.
Product parameters:
PID instrument - - reading + 20% or + 0.5ppm, large value, from 0.5ppm to 2000ppm. This accuracy is calibrated with isobutene at 100~300ppm (including drifting) on time temperature and humidity at school.
FID instrument - - reading + 10% or + 1.0ppm, large value, from 1.0ppm to 10000ppm. This accuracy is used for single point calibration of methane at 100~500ppm (including drifting) on time temperature and humidity at school. In the range of 3000 to 30000ppm, the calibration error of the calibration readings is + 10% with 15000ppm methane.
Repeatability
PID instrument - - 100ppm isobutylene when + 1%.
FID instrument - - 100ppm methane + 2%
dynamic range
PID instrument - - 0.5~2000ppm isobutene
FID instrument - - 1.0~30000ppm methane
Linear range
PID instrument - - 0.5~200ppm isobutene
FID instrument - - 1.0~30000ppm - methane
Minimum detection limit
The minimum detection limit is defined as 7 times the standard inter peak noise
PID instrument ----0.5ppm benzene
FID instrument ----0.5ppm ethane
Response time using a close range sampler
PID instrument - - 500ppm isobutene, reached the final value of 90% in 3.5 seconds at most
FID instrument - - 10000ppm methane, reach the final value of 90% in 3.5 seconds at most
Use of the recovery time of the close range sampler
PID instrument - - use 500ppm to return to baseline of isobutene, up to 10% in 4.5 seconds
FID instrument - - 10000ppm methane, most returned to baseline value of 10% in 4.5 seconds
Response time using an extended probe
PID instrument - - 500ppm isobutene, reached the final value of 90% in 15 seconds at most
FID instrument - - 10000PPm methane, reach the final value of 90% in 15 seconds at most
Recovery time using an extended probe
PID instrument - - use 500ppm to return to baseline of isobutene, up to 10% in 20 seconds
FID instrument - - 10000PPm methane, most returned to baseline value of 10% in 20 seconds
Use of the response time of the active carbon filter
PID instrument - - 100ppm isobutene, reached the final value of 90% in 20 seconds at most
FID instrument - - 10000PPm methane, reach the final value of 90% in 20 seconds at most
Use of the recovery time of the active carbon filter
PID instrument - - 100ppm up to isobutene, back to the original value of 10% in 20 seconds
FID instrument - - 10000PPm methane, most back to the original value of 10% in 20 seconds
Data storage interval
Auto Mode automatic mode - - 1 / sec to 1 /999, VOC or FE (user selectable volatile leakage) Mode mode - - - 2 to 30 seconds, the user selectable. The sampling rate is at the entrance of the sampling probe, and the rating is 1LPM..
Battery
When the battery is filled with electricity for up to 10 hours and at 20 degrees Celsius (32 degrees Fahrenheit), the working time is at least 10 hours. The use of background light on the probe will shorten the battery life. The battery can be replaced by opening the back of the instrument.
Warning: do not replace batteries in any dangerous area due to the presence of flammable gases
Battery charger
Battery charger is an independent unit that can recharge the battery inside the instrument while the instrument is working. The charger is full of electricity in the battery for up to 14 hours. If the instrument is charged when working, the charging time will be longer. Two types of charger are available, one is 90VAC alternating current, 50/60HZ or 230VAC AC, 50/60HZ.
(warning: because of the existence of combustible gases,
PID lamp life routine cleaning, lamp life over 2000 hours
FID life is regular cleaning, more than 5000 hours
The audio output level is more than 90dB within 3 feet.
Hydrogen bottle volume pressure -- at 25 DEG C up to 15.3Mpa (77 2200Psi) ----85ml bottles
(5.19in3)
The working time of the hydrogen gas cylinder starts from the cylinder pressure of 15.3Mpa (2200Psi) and works for more than 8 hours.
Shell description
The shell of the analyzer and the front panel are made of heat-resistant and chemically resistant plastic. Cleaner
The shell of the device (10% chlorine solution).
Electrical interface connection
The probe connection is waterproof.
Warning: do not use charger and USB interface in dangerous areas due to the existence of combustible gas.
The mechanical connection sampling line is a fast connecting device.
Portability
The instrument provides a detachable shoulder band.
Hold-all
Each instrument provides the necessary tools, including the special tools and tools to open the FID and PID packages.
A flow limiter for calibrating the flow of the pump.
Rough weight
Analyzer:
Only FID---4.8kg (9.2 lb)
PID+FID--- (9.4lb)
Probe component:
Standard probe -- - -- (0.5 lb)
Enhanced probe - (1.5 lb)